POTENTIAL INFLUENCES ON THE MEASUREMENT READINGS
Substance Interference from
Paracetamol > 18 mg/dL
Bilirubin > 20 mg/dL
Pralidoxime iodide > 200 mg/dL
Triacylglycerol > 1000 mg/dL
Cholesterol > 300 mg/dL
Glutathione > 74 mg/dL
Uric acid > 20.4 mg/dL
If haematocrit values are less than 20 %, it is possible that an
inaccurately high blood glucose value is displayed. If haematocrit
values are more than 70 %, it is possible that an inaccurately low
blood glucose value is displayed.
When measuring EDTA-treated blood samples with the iDia blood
glucose meter, the blood glucose values recorded may dier from
those of an untreated blood sample. Therefore, the use of sample
tubes treated with lithium-heparin is recommended.