IMI Sensors: A PCB Piezotronics Div 699A07 User Manual
MAN-0192 rev A Page 12 of 39 800-959-4464
1. Press FILE dial to select “Traceability.” A screen with the following information will
appear:
• Model
• Serial Number
• Firmware Revision Number
• Calibration Date
• Reference Sensor Sensitivity
• PRD-P
• NIST Traceability Number
• PTB Traceability Number
2. Press FILE dial to go back to main screen and shaker will go back to shaking.
Test Settings
The “Test Settings” menu can be found by pressing FREQUENCY dial > “Test
Settings.” A screen with the following will appear, use the FREQUENCY dial to
highlight and toggle all settings:
• Back – returns user to “Calibration Options” menu
• Cal Route: N/A, Active or off
o N/A indicates the Calibration Route firmware option has not been
purchased. Calibration Route allows users to program semi-automated test
points with instant pass/fail notification. See “Calibration Route” section for
more information. Contact IMI Sensors to unlock this feature.
• Source: Internal or External
o If external is selected the shaker can be controlled with an external source.
See “Input/Output” for more information.
• Frequency Unit: Hertz or CPM (cycles per minute)
• Sensor Type: ICP®, Voltage or Modulated Current
o Use ICP® for most accelerometers, Voltage for proximity probes and
moving coil sensors, Modulated Current for high-temp turbine vibration
sensors with AC current output. See “Selecting the Input Mode” for more
information. Modulated current is optional; contact IMI Sensors to unlock
this feature.
• Sensor Readout: mV/EU or mV
o Changes the display to show the sensor under test’s sensitivity (output
voltage divided by input vibration, mV/EU) or the sensor under test’s raw
AC output voltage.
Amplitude Units
Amplitude units that are seldom or never used can be turned off by using the
“Amplitude Units” feature, found by pressing FREQUENCY dial > “Amplitude Units.”
The “Amplitude Unit Screen” shows all 10 available amplitude scales on model
699A07 Portable Vibration Calibrator. Use the FREQUENCY dial to highlight each
scale and press the dial to toggle the scale on or off. A filled circle next to the scale
indicates it is active. An empty circle next to the scale indicates it is inactive. Inactive
scales do not appear when cycling through scales using the AMPLITUDE dial
during normal operation.