IMI Sensors: A PCB Piezotronics Div 699A07 User Manual
MAN-0192 rev A Page 11 of 39 800-959-4464
Route CreatorIf firmware option 600A31 is ordered, this tab can be used to
create semi-automated tests with instant pass/fail notification for almost any
vibration sensor. See “Calibration Route” for more information.
Additional Features
Delete
The “Delete” feature can be found under the FILE dial > “Delete.” When “Delete” is
selected, the shaker will stop moving and four options will appear:
1. “Delete Point” will delete a current point.
2. “Delete Record” will delete the entire current record.
3. “Delete All” will delete all data points and all records that are stored on the internal unit
memory.
4. “Back” will return to the main screen.
USB Options
The “USB Options” feature can be found under the FILE dial > “Tools” > “USB
Menu” > “USB Options.” When “USB Options” is selected, the following information
will appear on the screen:
“Status” USB flash drive connected or not connected.
“Partition”Format of USB flash drive connected to the unit.
“Available” Memory space available on USB flash drive.
“Required” Space required to save all records on USB flash drive.
And the following actions are available:
“Eject Drive” will safely eject the USB flash drive from the unit.
“Format USB” for formatting the USB flash drive. (FAT32 partition)
“Back” to go back to the USB Menu.
Date and Time
The “Date and Time” feature can be found under the FILE dial > Tools> Options
> “Date and Time.”
1. Press the FILE dial to select “Adjust.”
2. Turn the AMPLITUDE dial to the select the current month, day and year and push the
FREQUENCY dial to confirm or the AMPLITUDE dial to change.
3. Press the AMPLITUDE dial to select “yes” this is correct.
4. Turn the AMPLITUDE dial to select the current hours and minutes then push the
FREQUENCY dial to confirm the time is correct.
5. Press the AMPLITUDE dial to select “yes” this is correct.
Calibration Interval
The “Calibration Interval” can be adjusted under the FILE dial > “Tools” > “Options”
> “Calibration Interval.”
1. Select “Adjust” and press FILE dial.
2. Turn AMPLITUDE dial to select number of months for Calibration Interval.
Suggested Best Practice: 12 months. The calibration interval can be defined to be
anywhere from 1 to 72 months, or set up to never expire.
3. Press FREQUENCY dial to confirm selected Calibration Interval.
Traceability
The “Traceability” feature can be found under FILE dial > “Tools.”
69142
REV. A