ACF5 0601.01 GB
www.imo.se
17
Strainer
The pump has to be protected from foreign maers
such as weld slag, pipe scale, etc., that could enter the
pump via the suction line. If the cleanliness of the sys-
tem cannot be guaranteed, a strainer must be installed
in the inlet pipe near the pump. For practical reasons
a suction strainer with 0.8-2.0 mm mesh openings is
recommended.
The size of the strainer should be selected so that it is
large enough to allow adequate pressure at the pump
inlet. The pressure drop across the strainer should
preferably not exceed 0.1 bar at max. ow rate and nor-
mal operating viscosity. A vacuum gauge between the
strainer and the pump inlet is recommended to indicate
when the strainer needs cleaning.
17
m
Oil leakage may make the oor slippery
and cause personal injury
18
m
When handling liquids that may harm
skin, use gloves and/or protective clothing
19
m
When handling liquids which may involve
re hazard, appropriate precautions to
avoid danger are to be taken.
Fig. 4 Distance between coupling halves.
(IMO AB standard coupling)
Outer diameter
of coupling
(D mm)
Distance between
coupling halves
(t mm)
Outer diameter
of coupling
(D mm)
Distance between
coupling halves
(t mm)
A B A B
50 26 2.0 8 148 3.5
67 40 2.5 16 168 3.5
82 55 3.0 18 194 3.5
97 65 3.0 20 214 4.0
112 80 3.5 24 240 4.0
128 95 3.5 26
Fig. 3 Alignment of the IMO AB standard coupling
D4 ø max 0.3 mm
D6 ø max 0.4 mm
E4 ø max 0.4 mm
See table below
An angular
misalignment of
0.1° corresponds to
approx. 0.2 mm
deviation/100 mm.
Angular
alignment
Distance
between
coupling halves
Circular
run-out
A
x = y - t
B