CM480 - Check operating panel for general condition
Orientation
Check operating panel for general condition, keyboard, tightness
Procedure
1. Check operating panel for tightness – free of oil and condensate.
2. Check operating panel keyboard, and replace keys if necessary.
3. Record replacement date of the backup battery of the operating panel and the
ARTIS card.
Maintenance Summary - 2.000 Operating hours