tina03e1 (0206) BPG400 v1.om 19
In case of severe contamination the baffle can be easily replaced. For this purpose,
the gauge first has to be deinstalled (Deinstallation of the gauge ® 2 40).
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the
desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
· New baffle (® 2 47)
· Screwdriver No 1
· Pin (e.g. pencil)
Carefully remove the baffle with the screwdriver.
Place new baffle carefully onto the sensor opening.
3.1.3.2 Replacing the Baffle
Required tools / material
Procedure