Inficon BPG400 - Replacing the Sensor Unit

Inficon BPG400
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 tina03e1 (0206) BPG400 v1.om
Replacement is necessary, when
· the sensor is severely contaminated
· the sensor is mechanically deformed
· the sensor is faulty, e.g. filament of hot cathode broken (® 2 44)
· the sensor is faulty, e.g. Pirani element broken (® 2 44)
· Allen key, size 2.5 mm
· Spare sensor (® 2 47)
Deinstall the gauge (® 2 40).
Deinstall the electronics unit from the faulty sensor and mount it to the new
sensor (
® 2 15).
Adjust the gauge (® 2 42).
The new sensor is now installed.
6.4 Replacing the Sensor
Required tools / material
Procedure