EasyManua.ls Logo

Inficon BPG400 - Appendix B: Gas Type Dependence

Inficon BPG400
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50 tina03e1 (0206) BPG400 v1.om
Output signal U
[V] [mbar]
Pressure p
[Torr] [Pa]
0.3 / 0.5
Sensor error (
® 2 44)
0.51 … 0.774 Inadmissible range
0.774 5×10
-10
3.75×10
-10
5×10
-8
1.00 1×10
-9
7.5×10
-10
1×10
-7
1.75 1×10
-8
7.5×10
-9
1×10
-6
2.5 1×10
-7
7.5×10
-8
1×10
-5
3.25 1×10
-6
7.5×10
-7
1×10
-4
4.00 1×10
-5
7.5×10
-6
1×10
-3
4.75 1×10
-4
7.5×10
-5
1×10
-2
5.50 1×10
-3
7.5×10
-4
1×10
-1
6.25 1×10
-2
7.5×10
-3
1×10
0
7.00 1×10
-1
7.5×10
-2
1×10
1
7.75 1×10
0
7.5×10
-1
1×10
2
8.50 1×10
1
7.5×10
0
1×10
3
9.25 1×10
2
7.5×10
1
1×10
4
10.00 1×10
3
7.5×10
2
1×10
5
>10.00 Inadmissible range
Pressure indicated (gauge adjusted for air, Pirani-only mode)
10
2
8
6
4
2
10
1
8
6
4
2
10
0
8
6
4
2
10
–1
8
6
4
2
10
–2
8
6
4
2
10
–3
p (mbar)
p
eff
(mbar)
10
–3
246
10
–2
246
10
–1
246
10
0
246
10
1
246
10
2
H
2
He Ne
Xe
Ar
Freon 12
H
2
O
vapor
Air
O2
CO
N2
Kr
Indication range
above 10
-2
mbar)
CO
2
Conversion table
B: Gas Type Dependence
Indication range
above 10
-2
mbar

Table of Contents

Other manuals for Inficon BPG400