24 tina03e1 (0206) BPG400 v1.om
Plug the sensor connector into
the gauge and secure it with the
locking screws.
Connect the other end of the sensor cable to the connector of the instru-
ment or gauge controller you are using.
The gauge can now be operated via analog and RS232C interface.
For operating BPG400-SD via DeviceNet, an interface cable conforming to the
DeviceNet standard is required.
If no such cable is available, make one according to the following indications.
A shielded special 5 conductor cable conforming to the DeviceNet standard has to
be used (
® & [4], [6]).
Make the DeviceNet cable according to the following indications.
3
1
42
5
Micro-Style, 5 pins,
(DeviceNet)
female, soldering side
Pin Function (BPG400-SD)
1 Drain
2 Supply +24 VDC (DeviceNet) interface only
3 Supply common GND (DeviceNet interface only)
4 CAN_H
5 CAN_L
3.2.2.2 Making a DeviceNet
Interface Cable
(BPG400-SD)
Cable type
Procedure