6.3.4 Updating the software
Software updates from INFICON are installed with the aid of a USB flash drive. The
update function of the device can be found under "Functions > Data > Update".
An update is possible,
• if one or several updates are available on the USB flash drive, but only one update
per type at most (control unit, MSB box, I/O module),
• if these parts are also connected free of disturbances and have an update
function.
The corresponding buttons in the update menu such as "Control Unit", "MSB Box",
and "I/O Module" are active and can be activated individually.
NOTICE
Aborted connection
Data loss due to an aborted connection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
►
Switch the device off and back on after a software update has taken place.
6.3.4.1 Updating the software of the control unit
The software is included in two files named Handset_IFC_Vx.xx.xx.exe and
Handset_IFC_Vx.xx.xx.key.
1
Copy the file into the main directory of a USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
Select: "Functions > Data > Update > Control unit".
ð Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
4
Check the version information.
5
Select the "Start" button to start the update. Do not switch off the device and do
not remove the USB flash drive while the software is being updated!
6
Follow the instructions on the touchscreen and wait until the update is complete.
6.3.4.2 Checking and updating the software version of the MSB box
The current software is available from the Inficon support.
The functions of the XL Sniffer adapter set are taken into consideration in system
software version 2.11 or higher.