OPERATING MANUAL
SQC-310
Deposition Controller
PN 074-550-P1D
C
over
P
age
TM

Table of Contents

Questions and Answers

  • K
    Kimberly TorresAug 12, 2025
    How to troubleshoot if Inficon Controller SQC-310 does not turn on?
    • M
      Michael NixonAug 12, 2025
      If the Inficon Controller SQC-310 does not turn on, first, ensure the line cord is properly connected and the rear panel power switch is in the ON position. If that doesn't work, check that the line voltage is within the specified range for the SQC-310. Also, inspect the fuses in the power inlet and replace them if they are open with the specified fuse.
  • W
    Wendy AlvarezAug 16, 2025
    What to do if Inficon Controller SQC-310 locks up?
    • J
      jerry07Aug 16, 2025
      If the Inficon Controller SQC-310 “locks up”, make sure all covers and panels are securely fastened. Ensure that sensor, oscillator, source output, and I/O cables, along with the line cord, are at least 30.5 cm (1 ft.) away from high voltage/high power cables or any other sources of electrical noise. Check that ground wires or straps are short with a large surface area to minimize impedance to ground, and that they connect to an appropriate earth ground.
  • T
    Timothy MathewsAug 19, 2025
    Why does my Inficon Controller display 'Crystal Fail'?
    • S
      Shirley RodriguezAug 19, 2025
      If the crystal fail is displayed on your Inficon Controller, install a new crystal. If there are two crystals installed, remove the extra one, or reverse the crystal orientation. Inspect the crystal for scratches; if scratched, replace it. Ensure that the oscillator and sensor are connected to all active sensor channels set to On in the Edit Layer screen. Also, make sure that you are using a crystal with a starting frequency appropriate for the SQC-310 frequency range and change the Min/Max Frequency settings in the System Menu screen.
  • D
    Daniel HernandezAug 23, 2025
    How to fix noisy Rate, Thickness, and Frequency readings on Inficon SQC-310?
    • K
      Kimberly WilsonAug 23, 2025
      If the Rate, Thickness, and Frequency readings are noisy on your Inficon Controller, ensure that the in-vacuum cable length does not exceed 78.1 cm (30.75 in.) and use the 15.2 cm (6 in.) cable between the oscillator and feedthrough. Also, keep the sensor, oscillator cables, source output cables, I/O cables, communications cable, and line cord at least 30.5 cm (1 ft.) away from high voltage/high power cables or other sources of electrical noise. Check that ground wires or straps are short with a large surface area to minimize impedance to ground and connect to an appropriate earth ground.
  • C
    Charles EvansAug 27, 2025
    What to do if Inficon SQC-310 displays Crystal Fail during deposition?
    • M
      Michael MoralesAug 27, 2025
      If the crystal fail is displayed during deposition before the normal life of the crystal is exceeded on your Inficon Controller, use a shutter to shield the sensor during source conditioning and move the crystal farther away (at least 25.4 cm (10 in.)) from the source. Replace the crystal, and consider using an Alloy crystal if appropriate for the deposited material. Visually inspect the crystal for an uneven coating and correct the shutter or sensor positioning if needed. Also, Poor Rate control may be triggering Xtal Quality. Unstable/noisy crystal may be triggering Xtal Stability. Replace crystal. Correct any external conditions (e-beam arcing, thermal changes, etc.) triggering Xtal Stability. You can change the values, or disable Xtal Quality and/or Xtal Stability.
  • J
    Jamie AdamsAug 30, 2025
    Why is the frequency reading incorrect on my Inficon SQC-310 Controller?
    • J
      Jordan FisherAug 30, 2025
      If the frequency reading in the Sensor Information screen of your Inficon Controller is an incorrect value, ensure that the in-vacuum cable length does not exceed 78.1 cm (30.75 in.) and use the 15.2 cm (6 in.) cable between the oscillator and feedthrough.
  • N
    nguyenamberSep 3, 2025
    How to resolve large thickness reading excursions during deposition on Inficon SQC-310?
    • M
      martinezkarenSep 4, 2025
      If the thickness reading has large excursions during deposition with your Inficon Controller, replace the crystal. Ensure a straight line from the center of the source to the center of the crystal is perpendicular to the face of the crystal. Also, check water flow rate and temperature for sensor cooling.
  • M
    micheal31Sep 7, 2025
    Why is the frequency reading unstable on my Inficon SQC-310?
    • D
      Dennis StephensSep 7, 2025
      If the frequency reading in the Sensor Information screen of your Inficon Controller is unstable or drifting, control the vacuum chamber temperature. Move the crystal farther away from the source (at least 25.4 cm (10 in.) from source). Check sensor water cooling for correct flow and temperature. Avoid condensation by turning off cooling water to the sensor before opening the vacuum chamber to air, and then flow heated water above the dew point of the room through the sensor when the chamber is open. Replace crystal. Use an ohmmeter to check electrical continuity and isolation of cables.

Summary

Chapter 1: Introduction

1.2 SQC-310 Safety

Critical safety information, including definitions of terms.

1.2.2 General Safety Information

Essential safety precautions for operating the device.

1.2.3 Earth Ground

Specific requirements for proper earth grounding.

1.4 SQC-310 Specifications

Detailed technical specifications for the SQC-310.

Chapter 2: Installation

2.4 System Connections

Diagram and explanation of system connections.

2.5 Ground Requirements

Critical safety guidelines for proper system grounding.

2.5.1 Establishing Earth Ground

Detailed procedure for setting up an earth ground connection.

Chapter 3: Operation

3.2 SQC-310 Menu Overview

Introduction to the device's menu system.

3.3 Main Screen, Menu 1

Detailed explanation of Main Screen Menu 1 functions.

3.4 Main Screen, Menu 2

Detailed explanation of Main Screen Menu 2 functions.

3.8 Process Menu - Edit Layer

Detailed settings for configuring individual layers.

3.12 System Menu

Configuring system-wide parameters and settings.

Chapter 4: Communications

4.2 SQC-310 Communications Protocol

Description of the data communication protocol.

4.3 SQC-310 Commands

Comprehensive list of available commands for control and data retrieval.

Chapter 5: SQC-310 Comm Software

Chapter 6: Troubleshooting and Maintenance

6.1 Troubleshooting Guide

Comprehensive guide to diagnose and resolve common issues.

Chapter 7: Calibration Procedures

7.5 Control Loop

Explanation of PID control loop parameters and their function.

7.6 Control Loop Tuning Procedure

Detailed guide for tuning the PID control loop for stable deposition.

Inficon SQC-310 Specifications

General IconGeneral
ModelSQC-310
CategoryController
TypeThin Film Deposition Controller
Measurement TechnologyQuartz Crystal Microbalance (QCM)
Channels1
Input Voltage100-240 VAC
Operating Temperature15 to 35°C
Measurement PrincipleQuartz Crystal Microbalance (QCM)
Sensor CompatibilityInficon QCM Sensors
Power Requirements50/60 Hz
Communication InterfacesRS-232
Storage Temperature-20 to 70 °C