Why Inficon IC6 Controller displays 'CoDep Not Calibr(ated)'?
- WWilliam WoodsAug 20, 2025
If the co-deposited materials lack calibrated thickness values in the material setup for all active sensors, the Cross Talk calibration becomes uncalibrated. To resolve this, ensure that the Material parameter Cal Thickness values are entered for all active sensors, either manually or via automatic calibration. Alternatively, set the Layer parameter Co-Dep to No.