5 - 2
IPN 074-505-P1E
IC6 Operating Manual
#, move the cursor to the appropriate library display: Lib A-Hf, Lib Hf-Sc or Lib Sc-Z.
Press F4 to clear a Material of all programmed values. Move the cursor to a
Material to be cleared. Press the "Default Material" function key F4. This will set the
Material parameter values, including its name, to their dafault values.
5.1.2 Material Definition
Figure 5-2 Material Library A - Hf display
The three Material LibrariesLib A-Hf, Lib Hf-Sc and Lib Sc-Zprovide an
alphabetic list of materials by chemical name along with their density and Z-Ratio.
Once a material is selected from the library listing by moving the cursor to it, press
F1 Define Material. The correct chemical formula and library name, density and
Z-Ratio values will now be imported to this Material # and the Material/Source page
will appear. To rename this material, cursor to the name and use the keys in
cell-phone format to enter up to 15 characters.
To move among and through the materials, use the four cursor keys.