4 - 28
PN 074-550-P1C
SQC-310 Operating Manual
4.3.13.1 Get Output Rate Example
To get the output 1 rate:
Command: !$M1(92)(113)
Response: !)A 0.00(87)(100)
4.3.14 Get Sensor Thickness
Get Command Message = <CommandID><Sensor Number>
Get Response Message = <A><String|Value>
4.3.14.1 Get Sensor Thickness Example
Get Sensor 1 thickness:
Command: !$N1(93)(81)
Response: !*A 0.000(90)(92)
4.3.15 Get Output Thickness
Command Message = <CommandID><Output Number>
Response Message = <A><String|Value>
4.3.15.1 Get Output Thickness Example
Get Output 1 thickness:
Command: !$O1(103)(146)
Response: !*A 0.000(90)(92)
Table 4-15 Get sensor thickness parameters
Cmd ID Description Parameters
N Returns the thickness reading for
the requested sensor.
Sensor Number (1 to 4)
Table 4-16 Get output thickness parameters
Cmd ID Description Parameters
O Returns the average thickness of all
sensors assigned to the requested
output.
Output Number (1 to 4)