4 - 31
PN 074-550-P1C
SQC-310 Operating Manual
4.3.19.1 Set Active Process Example
Set the active process to Process 1:
Command: !$T1(104)(50)
Response: !$A5 (53)(151)
4.3.20 Set Run State
Set Command Message = <CommandID><Parameter>
Set Response Message = <A>
4.3.20.1 Set Run State Example
To start the Active process:
Command: !$U0(155)(82)
Response: !$A5 (53)(151)
Table 4-21 Set run state parameters
Cmd ID Description Parameters
U Sets SQC-310 operating state
0 Start Process
1 Stop Process
2 Start Layer
3 Stop Layer
4Next Layer
5 Force Final Thickness
32 Zero Thickness
33 Zero Time
38 Soak Hold Enable
39 Soak Hold Disable