4 - 27
PN 074-550-P1C
SQC-310 Operating Manual
4.3.11.1 Get Readings Examples
To Get Output readings:
Command: !$K1(93)(49)
Response: !cA0.00 0.00 100.00 0.000 0.00 -1 -1 -1 -1 -1 -1 -1 -1 -1 -1 -1 -1(44)(90)
The return string is of the form: Time, Rate 1, Deviation 1, Thickness 1, Power 1,
Rate 2, Deviation 2, Thickness 2, etc.
To Get Sensor readings:
Command: !$K2(157)(49)
Response: ! 80A0.00 0.00 0.000 5950000.00 0.00 0.000 5950000.00 0.00 0.000
5950000.00 0.00 0.000 5950000.00(80)(85)
The return string is of the form: Time, Rate 1, Thickness 1, Frequency 1, Rate 2,
Thickness 2, etc.
4.3.12 Get Sensor Rate
Get Command Message = <CommandID><Sensor Number>
Get Response Message = <A><String|Value>
4.3.12.1 Get Sensor Rate Example
To get the sensor 1 rate:
Command: !$L1(102)(50)
Response: !)A 0.00(87)(100)
4.3.13 Get Output Rate
Get Command Message = <CommandID><Output Number>
Get Response Message = <A><String|Value>
Table 4-13 Get sensor rate parameters
Cmd ID Description Parameters
L Returns the sensor rate for the
requested sensor.
Sensor Number (1 to 4)
Table 4-14 Get output rate parameters
Cmd ID Description Parameters
M Returns the average rate of all
sensors assigned to the requested
output.
Output Number (1 to 4)