4 - 19
PN 074-550-P1C
SQC-310 Operating Manual
4.3.5.1 Get/Set Layer Parameter Examples
Get Process 1, Layer 1, Source 1 initial rate and final thickness:
Command: !.DA1 1 1? 1 2(90)(89)
Response: !,A1,0 2,0 V (86)(133)
Get layer status information:
Command: !-DB? 1 2 3 4(40)(67)
Response: !FA1,4:4:4 2,1 3,1.1:1.1:1.1 4,1:1:1 9(154)(60)
To Delete all layers:
Command: !#E(143)(54)
Response: !$A5 (53)(151)
4.3.6 Get/Set Material Parameters
Only one material may be Get/Set at a time.
Get Command = <CommandID><Material Number><?><Space><Parameter>
Get Response = <A><String|Value>
Set Command = <CommandID><Material Number><Space><Parameter>
<Comma><String|Value>
Set Response = <A>
DB Gets the layer that is currently
running or set to run next if not
currently running a layer.
Message format changes to: DB?
[Parameter]
1 Current layer - Layer number
2 Current layer - NonCoDep Position
3 Current layer - NonCoDep Position(s)
with Source Number
4 Current layer - CoDep Positions
E Deletes all 1000 layers. Most often
used to clear SQC-310 in
preparation for downloading a new
list.
None
Table 4-6 Get/Set layer parameters (continued)
Cmd ID Description Parameters
Table 4-7 Get/Set material parameters
Cmd ID Description Parameters
F Gets/Sets the parameters of the
100 stored materials.
1 Material Name
2Density
3 Z-Factor (Z-Ratio)