4 - 24
PN 074-550-P1C
SQC-310 Operating Manual
4.3.8.1 Get/Set Sensor and Source Parameter Examples
Get sensor 1 number of positions, control type, and current crystal position:
Command: !,HA1? 2 5 9(153)(140)
Response: !0A2,1 5,1 9,1 (67)(113)
Get source 1 number of positions, control type, and current pocket position:
Command: !,HB1? 2 5 9(77)(83)
Response: !0A2,1 5,3 9,1 8(142)(52)
Get status flags for sensor 1:
Command: !*HC1? 5 6(50)(112)
Response: !,A5,0 6,0 e (101)(149)
Get status flags (in use) for source 1:
Command: !(HD1? 1(86)(130)
Response: !(A1,1 (58)(116)
Get I/O mapping of sensor 1:
Command: !*HE1? 1 2(63)(37)
Response: !hA1,0:0:0:0:0:0:0:0:0:0:0:0:0:0:0:0 2,0:0:0:0:0:0:0:0:0:0:0:0:0:0:0:0 [
(91)(147)
Get I/O mapping of source 2:
Command: !*HF1? 1 2(38)(82)
Response: hA1,0:0:0:0:0:0:0:0:0:0:0:0:0:0:0:0 2,0:0:0:0:0:0:0:0:0:0:0:0:0:0:0:0 [
(91)(147)