4.3.6.1 Get/Set Material Parameter Examples
Get Material 1 name:
Command: !'F1? 1(135)(105)
Response: !,AAluminum9 (57)(152)
Set Material 1 name:
Command: !/F1 1,Aluminum(99)(39)
Response: !$A5 (53)(151)
4.3.7 Get/Set Input and Relay Parameters
Get Command = <CommandID><Input#|Relay#><?><Space> <Parameter>
Get Response = <A><Parameter><Comma> <String|Value>
Set Command = <CommandID><Input#|Relay#><Space><Parameter>
<Comma> <String|Value>
Set Response = <A>
Table 4-8 Get/Set input and relay parameters
Cmd ID Description Parameters
G Gets/Sets the logical function of
each of the 8 or 16 digital inputs.
1 Start Process
2 Abort Process
3 Start Layer
4 Stop Layer
5 Start Next Layer
6 Force Final Thickness
7 to 31 Start Process 1 to 25
32 Soak Hold
33 Zero Thickness
34 Zero Time
35 Out1 Pocket Ready
36 Out2 Pocket Ready
37 Out3 Pocket Ready
38 Out4 Pocket Ready
GA Gets/Sets the parameters of each
of the 8 or 16 digital inputs.
When changing the name (GA1),
the Name Mode command (GA5)
must be sent as the next
parameter.
1Name
2 Active Level
3 Pin Number
4 Input in Use
5 Name Mode
GB Gets/Sets the parameters of each
of the 8 or 16 digital relays.
When changing the name (GB1),
the Name Mode command (GB5)
must be sent as the next
parameter.
1Name
2 Type (N.O or N.C)
3 Pulses
4 Pulse Width (ms)
5 Pin Number
6 Relay in Use
7 Name Mode