28
tira85e1-a (2014-10) PxG55x.cp
High Trip Point
Time t
in-
active
activein-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
Index SI DataType NV Access PM Name
800E 0x01 BOOL x RW High Trip Enable
0x02 BOOL x RW Low Trip Enable
0x11 REAL x RW High Trip Point Limit
0x12 UDINT x RW High Trip Source Index
0x14 REAL x RW Low Trip Point Limit
0x15 UDINT x RW Low Trip Source Index
0x17 REAL x RW High Trip Hysteresis
0x18 REAL x RW Low Trip Hysteresis
0x1A UDINT x RW Source Index
High Trip Point
0
1
Disable
Enable
Low Trip Point
0
1
Disable
Enable
High Trip Point Limit: High limit to trigger trip point condition if Input Value
(I 0x900E, SI 0x01) is above this limit.
Object index of High Trip Point Value source.
Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved
For a fix limit set value to 0x800E1100 (High Trip Point Limit).
Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900E,
SI 0x01) is below this limit.
Object index of Low Trip Point Value source.
Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved
For a fix limit set value to 0x800E1400.
Subindex 0x01
Subindex 0x02
Subindex 0x11
Subindex 0x12
Subindex 0x14
Subindex 0x15