tira85e1-a (2014-10) PxG55x.cp 57
Execution of this command will store all parameters to non-volatile memory. If a
device automatically saves all non-volatile parameters at the time they are set, this
command will not take any action.
Index SI DataType NV Access PM Name
FBF2 0x01 V_STRING(4) RW Command
0x02 BYTE RO Status
0x03 V_STRING(1) RO Response
A device reset is initiated when the following byte sequence is sent.
Command
Byte 0
Byte 1
Byte 2
Byte 3
0x65
0x76
0x61
0x73
Status (supported values)
0
1
2
3
255
Last command completed, no error, no response
Reserved
Last command completed, error, no response
Reserved
Command is executing
Response
Byte 0 See Subindex 0x02
4.19.5 Store Parameters
Command
Subindex 0x01
Subindex 0x02
Subindex 0x03