[ 27 ]
EasyGluco
TM
Troubleshooting
Please refer to the table below to identify problems that persist with no messages appearing.
PROBLEM CAUSED BY WHAT TO DO
The meter does not
power on
There may be a problem with
the test strip
Re-test with a new strip.
Re-code the meter and re-test.
The codes of the meter and
of the test strip vial do not
match
Change the batteries and if the problem persists, contact
your local representative.
Batteries are dead or there is
a problem with the meter.
Poor amount of sample Apply a sufficient amount of the blood sample to a new test
strip and re-test.
Re-test with a new strip.
Problem with the test strip
Test does not start
after applying blood
sample
The result is doubtful
INFOPIA [ 28 ]
Specifications
Model IGM-0002A
Test Range 0.6~33.3 mmol/L (10~600 mg/dL)
Reading Time 9 sec
Memory Capacity 200 test results
Operating Temperature 50-104˚F(10-40˚C )
Operating Humidity 10 ~ 90%
Sample Type Capillary Whole Blood
Calibration Plasma-Equivalent
Hematocrit 30 ~ 55%
Sample Volume Minimum 1.5 ㎕
Display Type LCD
Size (D X W X H) 22.8
X 54.5 X 91 (mm)
Weight 55g (including batteries)
Power Source 3V Li Battery (CR2032)
X 2
Battery Life running 5000 tests
Glucose Unit Either mmol/L or mg/dL
Specifications