20
www.infopia21.com
Click
1. Unscrew the tip of the lancing
device by turning it counter-
clockwise while holding the
base firmly.
4. Twist the cap of the lancet off.
2. With two fingers, pull out the
lancet carrier and hold it in
place.
5. Put the tip of the lancing
device back on and turn it
clockwise.
6. Hold the tip firmly in one hand
then pull out the sliding barrel
with the other hand. This will
cock the lancing device.
7. Place the lancing device in
place. Hold it firmly against
the side of the finger, with
the cap resting on the finger.
(The harder it is pressed the
deeper the puncture.) Press
the release button to take a
sample.
8. Unscrew the tip of the
lancing device by turning it
counter-clockwise.
9. Push the lancet ejector
forward with the thumb
and simultaneously pull
out the sliding barrel to
dispose of the used lancet
in a proper biohazard
container.
3. While holding the lancet
carrier, insert a new sterile
lancet into the bottom of
lancet carrier.
Collecting a Drop of Blood
21
RECOMMENDED AVAILABLE LANCETS
Some general sterilized lancets are BD-Ultra Fine, Cleanlet Fine, E-Z Ject, G-P Lite, Microlet, Monolet,
One Touch, Soft Touch, Therasense, Ultra TLC and Unilet GP.
Note: Some less known lancets may not work properly with the due to their dimensions. Please, contact
your supplier to make sure you are using proper lancets.
USING THE CLEAR TIP (AST)
1. Place the lancing device on the skin in the chosen area, press and hold it continuously for a few
seconds. Then, push the release button to take a sample.
2. Watch through the clear tip until a sufficient blood sample is taken. If there is not enough blood, gently
massage the area until a sufficient sample has been collected.
STORAGE
Products must be stored at room temperature protected from sunlight or moisture.
CLEANING AND DISINFECTION
Clean and wipe the outside of the lancing device a minimum of once a week with a soft cloth, dampened
with mild soap and water. Wash the adjustable tip and the clear tip once a week with mild soap. DO NOT
immerse the lancing device in liquid.
WARRANTY
Lancing device has a 2-year warranty from the date of purchase. A malfunctioning device should be
returned to the following address:
GMMC
Room No. 1112, Ace Tower 9th Bldg., 345-30, Kasan-Dong, Kumchon-Ku, Seoul, South Korea
Tel : +82 2 2104 0470
Fax : +82 2 2104 0472
GMMC, S.L.
C/Jordi de S. Jordi 13, 6-23 46022 Valencia SPAIN
Tel : +34 963 691 484
Fax: +34 961 125 949
Collecting a Drop of Blood
2. Pull the lancet out and discard accordingly.
1. Push the needle into the protective cover.
Disposing of your lancet