8
Connecting to Compatible Oxygen Concentrators
Note: Only connect the TAV® system to com-
patible oxygen concentrators that have the
TAV® logo (shown below).
1. Attach one end of the supply tubing to
the connector on the bottom right of
the device.
2. Attach the other end of the supply
tubing to a TAV® compatible oxygen
concentrator. Make sure that both ends
of the supply tubing are rmly pressed
onto each connector.
Supply
Tubing
TA
Logo
3. Turn the ow meter knob max counter-
clockwise when connected to the TAV.
See the concentrator’s Instructions For Use
(IFU) for operating instructions.
Note: To ensure proper operation, the
TAV® device must be connected to a TA
compatible oxygen concentrators.
Compatible concentrators are labeled with
the TAV® logo Turn the ow meter to Max
when the TAV device is connected or the
system will not operate properly. The TAV
device will not operate properly when
connected to concentrators without the TA
logo. Contact Inogen or visit
www.TidalAssist.com for a list of compatible
concentrators.