Technical Specications
Product Name: Tidal Assist Ventilator (TAV) System
Model No: VC-100 (Controller), RP-1000 (Nasal Pillow Interface)
Electrical:
1.5 VDC (one alkaline battery, Type AA/LR6)
Note: The use of other batteries (like “heavy duty or 1.5V rechargeable cells) will not
damage the system, but the battery life and accuracy of the low battery indication may be
aected.
Battery Life:
Approximately 90 days (at 8 hours of use/day)
Operating Temperature/Humidity:
41°F ~ 104°F (5°C ~ 40°C)
15% ~ 93% RH non-condensing
700 ~ 1060 hPa
Storage Temperature/Humidity/Air Pressure:
-13°F ~ 158°F (-25°C ~ 70°C)
10% ~ 93% RH non-condensing
700 ~ 1060 hPa (-1,000 to ~10,000 feet)
Enclosure Rating:
IP22 (ingress protection)
IP22 means that the TAV® device is designed to not allow touching of any hazardous parts
of the device for objects > 12.5mm (e.g., ngers). In addition, the TAV® enclosure provides
protection against dripping water up to an angle of 15 degrees.
Alarm Volume:
65 - 80 dBA (measured at 1 m)
Weight and Dimensions:
Approximately 3.8 ounces with battery (108 grams)
Approximately 5.25 (W) x 2.5 (H) x 0.9 (D) inches
Predicted Life:
3 years for TAV® device
30 days for Nasal Pillow Interface (after use has begun)
Product Contents:
TAV® Device
Nasal Pillow Interface
Regulator Adaptor
AA battery
Instructions for Use and Quick Reference Guide
Warranty Card
Trigger sensitivity
-0.054 cmH2O
Note: The TAV® System and battery contain
electronic waste that requires appropriate
disposal according to your local ordinances.