4.2. Step of calibration operation
According to the second method which can enter setting menu,
C01-C39
step
Method of operation
display Remark
1 [C01 ] After you enter calibration mode,
it display [C01 ]
2 press
[C01 1] Weight unit
option1=kg
2=lb
3 press
press
press or
[C02 ]
[C02 0]
[C02 2]
Set decimal digits
option0/1/2/3/4
Select decimal digit
example two decimal point
[C02 2]
4 press
press
press or
[C03 ]
[C03 1]
[C03 5]
Set graduation
option1/2/5/10/20/50
Select required graduation
examplegraduation 5[C03 5]
5 press
press
press or /
[C04 ]
[0100.00]
[0100.00]
Max capacity
examplemax weighing 100kg:
[0100.00]
6
press
press
press
press
[C05 ]
[C05 0]
[C05 1]
[CAL 9]
[0000.00]
Zero calibration
Option
0=no need zero calibration
1=need zero calibration
calibration zero please choose 1 and
ensure scale is empty and “stable”
light is on
Ensure zero calibration, countdown.
Till show[0.00](example for two
decimal point)