EasyManuals Logo

Instrutech Super Bee CVM201 User Manual

Instrutech Super Bee CVM201
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
Instruction Manual CVM201 Super Bee
InstruTech, Inc. Page 17
6.2 Display - mbar
The table below shows the displayed readings at various pressures for selected gases when engineering units
selected is in mbar.
Displayed Pressure Readings vs. True Pressure for selected gases - Engineering units in mbar
True Pressure
N
2
Ar
He
O
2
CO
2
KR
Freon12
Freon22
D
2
Ne
CH
4
0 mbar
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
.0001 mbar
.0001
.0001
.0001
.0001
.0001
.0001
.0001
.0001
.0001
.0001
.0001
.0003 mbar
.0003
.0003
.0003
.0003
.0003
.0003
.0003
.0003
.0003
.0003
.0003
.0006 mbar
.0006
.0006
.0006
.0006
.0006
.0004
.0006
.0006
.0006
.0006
.0006
.0013 mbar
.0013
.0009
.0011
.0013
.0015
.0005
.0020
.0020
.0017
.0009
.0023
.0027 mbar
.0027
.0019
.0021
.0027
.0031
.0013
.0041
.0041
.0032
.0020
.0044
.0067 mbar
.0067
.0044
.0053
.0067
.0059
.0031
.0101
.0093
.0080
.0047
.0102
.0133 mbar
.0133
.0088
.0107
.0129
.0146
.0064
.0195
.0179
.0161
.0095
.0203
.0260 mbar
.0260
.0174
.0214
.0263
.0295
.0126
.0398
.0362
.0323
.0187
.0405
.0666 mbar
.0666
.0431
.0539
.0655
.0731
.0313
.0966
.0919
.0799
.0463
0.100
0.130 mbar
0.130
.0857
0.110
0.120
0.140
.0623
0.190
0.180
0.160
0.100
0.210
0.260 mbar
0.260
0.160
0.210
0.250
0.270
0.120
0.360
0.340
0.330
0.180
0.410
0.666 mbar
0.666
0.410
0.570
0.640
0.650
0.280
0.810
0.790
0.91
0.470
1.04
1.33 mbar
1.33
0.790
1.25
1.29
1.26
0.530
1.39
1.38
2.06
0.990
2.13
2.66 mbar
2.66
1.51
2.95
2.58
2.27
0.930
2.15
2.21
5.50
2.11
4.43
6.66 mbar
6.66
3.26
17.9
6.63
4.45
1.70
3.26
3.49
327
6.98
10.0
13.3 mbar
13.3
5.33
OP
13.7
6.62
2.37
3.94
4.51
OP
28.6
37.1
26.6 mbar
26.6
7.73
OP
29.7
8.78
3.05
4.42
4.95
OP
778
473
66.6 mbar
66.6
10.4
OP
103
10.9
3.42
5.05
5.51
OP
OP
1012
133 mbar
133
11.7
OP
278
12.3
3.65
6.23
6.54
OP
OP
OP
266 mbar
266
13.0
OP
393
16.3
4.42
7.98
8.55
OP
OP
OP
400 mbar
400
15.0
OP
506
22.5
4.78
9.18
10.0
OP
OP
OP
533 mbar
533
17.9
OP
646
29.8
5.25
10.1
11.2
OP
OP
OP
666 mbar
666
21.4
OP
805
38.2
5.61
11.0
12.2
OP
OP
OP
800 mbar
800
25.0
OP
973
48.5
5.91
11.8
13.2
OP
OP
OP
933 mbar
933
29.0
OP
1140
61.4
6.19
12.5
14.2
OP
OP
OP
1011 mbar
1011
31.5
OP
1250
71.8
6.33
13.0
14.7
OP
OP
OP
1060 mbar
1060
33.4
OP
1320
79.1
6.45
13.2
15.1
OP
OP
OP
1190 mbar
1019
37.9
OP
OP
105
6.65
13.9
16.0
OP
OP
OP
1330 mbar
1330
43.3
OP
OP
147
6.77
14.7
16.9
OP
OP
OP
Values listed under each gas type are in mbar.
Notes:
1) OP = Overpressure indication; display will read “overpressure”.
Examples:
1) Gas used is nitrogen. Display shows pressure measurement of 13.3 mbar. True pressure of nitrogen is 13.3 mbar.
2) Gas used is argon. Display shows pressure measurement of 11.7 mbar. True pressure of argon is 133 mbar.
3) Gas used is CO
2
. Display shows pressure measurement of .0731 mbar. True pressure of CO
2
is .0666 mbar.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instrutech Super Bee CVM201 and is the answer not in the manual?

Instrutech Super Bee CVM201 Specifications

General IconGeneral
BrandInstrutech
ModelSuper Bee CVM201
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals