M L X 3 0 0 W A T T X E N O N L I G H T S O U R C E
ELECTROMAGNETIC COMPATIbILITY (EMC) USER INfORMATION
WARNING: Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying
documents.
WARNING: Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
WARNING: The equipment or system should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the
congurationinwhichitwillbeused.
NOTE: The EMC tables and other guidelines that are included in the Instruction Manual provide information to the
customer or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic
Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to
perform its intended use without disturbing other Equipment and Systems or non-medical electrical equipment.
Table 201
GuidanceandManufacturer’sDeclaration–EmissionsAllEquipmentandSystems
The MLX light source is intended for use in the electromagnetic environment specified below.
The customer or user of the MLX light source should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Enforcement – guidance
RF Emissions
CISPR 11
Group 1
The MLX unit uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
Radiated and
Conducted
Emissions
The MLX unit is suitable for use in all establishments including domestic, and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Conducted Emissions Tests Performed at both 240VAC, 50Hz and 120VAC,
60Hz
Harmonics
IEC 61000-3-2
N/A Equipment intended for Professional Use Only
Flicker
IEC 61000-3-3
N/A Equipment intended for Professional Use Only