14
M L X 3 0 0 W A T T X E N O N L I G H T S O U R C E
ELECTROMAGNETIC COMPATIbILITY (EMC) USER INfORMATION
Table 202
GuidanceandManufacturer’sDeclaration—ImmunityAllEquipmentandSystems
The MLX light source is intended for use in the electromagnetic environment specified below. The customer or user of
the MLX light source should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile. If
floors are synthetic, the relative humidity should be
at least 30%.
Electrical Fast
Transient/burst
IEC 61000-4-4
±2kV on AC Mains ±2kV on AC Mains Mains power quality should be that of a typical
commercial or hospital environment.
Note - Tests Performed at both 240VAC, 50Hz and
120VAC, 60Hz
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical
commercial or hospital environment.
Note - Tests Performed at both 240VAC, 50Hz and
120VAC, 60Hz
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the MLX unit requires continued operation during
power mains interruptions, it is recommended that
the MLX unit be powered from an uninterruptible
power supply or battery.
Note - Tests Performed at both 240VAC, 50Hz and
120VAC, 60Hz
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic fields should be that of a
typical location in a typical commercial or hospital
environment.
Note - Tests Performed at both 50Hz and 60Hz