M L X 3 0 0 W A T T X E N O N L I G H T S O U R C E
ELECTROMAGNETIC COMPATIbILITY (EMC) USER INfORMATION
Table 204
GuidanceandManufacturer’sDeclaration–EmissionsEquipmentandSystemsthatareNOT Life-Supporting
The MLX light source is intended for use in the electromagnetic environment specified below. The customer or user
of the MLX light source unit should ensure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to
80 MHz
V1 = 3 Vrms Portable and mobile RF communications equipment should
be separated from the MLX light source by no less than the
recommended separation distances calculated/listed below:
PVD )1/5.3(=
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
E1 = 3V/m
PED )1/5.3(=
80 to 800 MHz
PED )1/7(=
800 MHz to 2.5 GHz
Where P is the maximum power rating in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as determined by an
electromagnetic site survey, should be less that the compliance
levels (V1 and E1).
Interference may occur in the vicinity of equipment containing a
transmitter.
Table 206
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the MLX Equipment and Systems that are NOT Life-Supporting
The MLX light source is intended for use in the electromagnetic environment in which radiated disturbances
are controlled. The customer or user of the MLX light source can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF Communications Equipment and the MLX
light source as recommended below, according to the maximum output power of the communications
equipment.
Maximum
Output Power
(Watts)
Recommended Separation Distances for the MLX (meters)
150 kHz to 80
MHz
Pd 1667.1=
80 to 800MHz
Pd 1667.1
=
800 MHz to 2.5 GHz
Pd 3333.2
=
0.01 0.11667 0.11667 0.23333
0.1 0.36894 0.36894 0.73785
1 1.1667 1.1667 2.3333
10 3.6894 3.6894 7.3785
100 11.667 11.667 23.333

Questions and Answers: