Off time
2 – 7 sec.
A fixed set time, in which there is no tone.
Random off time
0 7 sec.
An extra time off, added to the ”Off time”, making it more
difficult for the patient to figure out when the tone is
present. The extra random time will vary throughout the
test randomly.
Example: If the Off time is set to 7 sec. and the random
time off is also set to 7 sec, the time off will vary between 7
and 14 sec.
Tone length
0.3 2 sec.
The duration of the tone given from the earphone.
Response window
2 9 sec.
The duration of time the patient must answer.