Luna - Instructions for Use - US Page 27
Off time
2 7 sec.
Time with no tone, recommended setting: 0.
Random off time
0 7 sec.
An extra time off, added to the ”Off time”, making it more
difficult for the patient to figure out when the tone is
present. The extra random time will vary throughout the
test randomly.
Example: If the Off time is set to 7 sec. and the random
time off is also set to 7 sec, the time off will vary between 7
and 14 sec.
Tone length
0.3 2 sec.
The duration of the tone given from the earphone.
Response window
2 9 sec.
The duration of time the patient has to answer.
X dB min. level
X dB start level
X dB maximum level
The lowest intensity the system will test to. If this is set for
10 dB, no frequencies will be tested below 10 dB
Will start the test tone at this level for each frequency
Your dB will never exceed X dB.
Test frequencies
125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000,
8000 Hz
Start frequencies
The frequency you want your test to begin with.