D-0128842-E – 2023/10
Luna - Instructions for Use - US Page 40
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Instrument (Luna) is intended for use in the electromagnetic environment specified below. The customer or the user of the Instrument
should assure that it is used in such an environment.
Electromagnetic environment - guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be
greater than 30%.
Immunity to proximity fields
from RF wireless
communications equipment
Spot freq. 385-5.785 MHz
Levels and modulation
defined in table 9
As defined in table 9
RF wireless communications equipment
should not be used close to any parts of the
Instrument.
transient/burst
+2 kV for power supply lines
+1 kV for input/output lines
Not applicable
+1 kV for input/output lines
Mains power quality should be that of a
typical commercial or residential environment.
Surge
IEC 61000-4-5
+1 kV Line to line
+2 kV Line to earth
Not applicable
Mains power quality should be that of a
typical commercial or residential environment.
Voltage dips, short
interruptions and voltage
variations on power supply
lines
IEC 61000-4-11
for 0.5 cycle, @ 0, 45, 90,
135, 180, 225, 270 and 315°
0% UT (100% dip in UT)
for 1 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
0% UT (100% dip in UT)
Not applicable
Mains power quality should be that of a
typical commercial or residential environment.
If the user of the Instrument requires
continued operation during power mains
interruptions, it is recommended that the
Instrument be powered from an
uninterruptable power supply or its battery.
(50/60 Hz)
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or residential environment.
Radiated fields in close
proximity — Immunity test
IEC 61000-4-39
9 kHz to 13.56 MHz.
Frequency, level and
modulation defined in
AMD 1: 2020, table 11
As defined in table 11
of AMD 1: 2020
If the Instrument contains magnetically
sensitive components or circuits, the
proximity magnetic fields should be no higher
than the test levels specified in Table 11
Note: UT is the A.C. mains voltage prior to application of the test level.