D-0114042-D – 2020/03
Sera™ - Additional Information Page 26
DPOAE protocols
The Sera
TM
device is delivered with one default DPOAE protocol (D00 2-5kHz, 3_4, SNR 6 dB). The
protocol on the device can be changed and additional protocols can be added to Sera
TM
through the
HearSIM
TM
software.
The available protocols are:
D00 2-5kHz, 3_4, SNR 6 dB
2
Stimulus level (L
1
/L
2
):
Stimulus tolerance:
Ratio f
2
/f
1:
Noise rejection level:
Max test time:
DP detection criteria
Min DP level:
SNR:
DP reliability:
DP stability:
# frequencies for a pass:
65/55 dB SPL
±7 dB
1.22
30 dB SPL
60 seconds
-5 dB SPL
6 dB
98.3%
±2 dB
3 / 4
D01 1.5-4kHz, 3_4, SNR 6 dB
2
Stimulus level (L1/L2):
Stimulus tolerance:
Ratio f
2
/f
1:
Noise rejection level:
Max test time:
DP detection criteria
Min DP level:
SNR:
DP reliability:
DP stability:
# frequencies for a pass:
65/55 dB SPL
±7 dB
1.22
30 dB SPL
60 seconds
-5 dB SPL
6 dB
98.3%
±2 dB
3 / 4
D02 1.5-6kHz, 3_5, SNR 6 dB
2
Stimulus level (L1/L2):
Stimulus tolerance:
Ratio f
2
/f
1:
Noise rejection level:
Max test time:
DP detection criteria
Min DP level:
SNR:
DP reliability:
DP stability:
# frequencies for a pass:
65/55 dB SPL
±7 dB
1.22
30 dB SPL
60 seconds
-5 dB SPL
6 dB
98.3%
±2 dB
3 / 5
D03 2-5 kHz, 65_55 dB SPL,
IEC
7
2
):
Stimulus level (L1/L2):
2, 3, 4, 5 kHz
65/55 dB SPL