EasyManua.ls Logo

Interacoustics Sera - Home screen; Adding a new patient

Interacoustics Sera
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0114042-D – 2020/03
Sera™ - Additional Information Page 30
TEOAE protocols
The Sera
TM
device is delivered with one default TEOAE protocol (T00 1.5-4kHz, 3_4, SNR 4 dB). The
protocol on the device can be changed and additional protocols can be added to Sera
TM
through the
HearSIM
TM
software.
The available protocols are:
Protocol name
Parameters
Algorithmic
sensitivity
T00 1.5-4 kHz, 3_4, SNR 4 dB
Center frequency bands:
Stimulus type:
Stimulus level:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
1.4, 2, 2.8, 4 kHz
Non-linear click
83 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
3 / 4
> 99.6%
T01 1.5-4 kHz, 2_4, SNR 6 dB
Center frequency bands:
Stimulus type:
Stimulus level:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
1.4, 2, 2.8, 4 kHz
Non-linear click
83 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
2 / 4
> 99.6%
T02 1.5-4 kHz, 3_4, 80 dB
SPL
Center frequency bands:
Stimulus type:
Stimulus intensity:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
1.4, 2, 2.8, 4 kHz
Non-linear click
80 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
3 / 4
> 99.6%