D-0114042-D – 2020/03
Sera™ - Additional Information Page 43
“Off Levels
message during
OAE test.
Probe is coming out of baby’s
ear or placement of probe has
otherwise changed during the
test so that stimulus levels are
outside of the acceptable
range.
Check the fit of probe in the baby’s ear. It may be
necessary to stop the test and start again after
securing the probe in the baby’s ear canal.