Invacare®CarrollCSSeries™
GuidanceandManufacturer’sDeclaration-ElectromagneticEmissions
EmissionsTest
Compliance
ElectromagneticEnvironment
-Guidance
RFemissions
CISPR11
ClassB
HarmonicemissionsIEC61000-3-2ClassA
Voltagefluctuations/flickeremissions
IEC61000-3-3
Complies
TheCSBedissuitableforuseinall
establishments,includingdomestic
establishmentsandthosedirectlyconnected
tothepubliclow-voltagepowersupply
networkthatsuppliesbuildingsused
fordomesticpurposes.
GuidanceandManufacturer’sDeclaration-ElectromagneticImmunity
TheCSBedisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheCSBedshouldassure
thatitisusedinsuchanenvironment.
ImmunityTest
IEC60601TestLevelComplianceLevel
Electromagnetic
Environment–Guidance
Electrostaticdischarge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concrete
orceramictile.Iffloorsare
coveredwithsyntheticmaterial,
therelativehumidityshouldbeat
least30%.
Electricalfasttransient/burst
IEC61000-4-4
±2kVforpowersupplylines
±1kVforinput/outputlines
±2kVforpowersupplylines
±1kVforinput/outputlines
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.
Surge
IEC61000-4-5
±1kVdifferentialmode
±2kVcommonmode
±1kVdifferentialmode
±2kVcommonmode
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.