What to do if the actuator is not working on Invacare Carroll CS3?
- OomcintyreAug 11, 2025
If the actuator on your Invacare Medical Equipment isn't working, there are several things you can check: * **System Fault:** Try resetting the system by holding the bed up and bed down buttons on the pendant for 5+ seconds. The bed will beep and stop when the reset is complete. Then, press the bed up button. If it beeps, the actuators are out of sync; raise the bed to the highest position to synchronize them. * **HI/LO Lock-out:** Make sure the lockout dial is in the unlocked position (horizontal) or the switch is set to OFF. * **Attendant Control Pad (ACP):** Ensure the lockout LED on the ACP is OFF by toggling the individual function lockout. * **Wire Connections:** Check for and connect any loose cable couplings or power supplies.