Troubleshooting
8Troubleshooting
8.1Troubleshooting
DANGER!
RiskOfInjuryOrDeath
Toreducetheriskofelectricalshock:
–Alwaysunplugthebedfromtheelectricaloutlet
beforeperformingmaintenance.
–DONOTunplugthepowercordfromthecontrol
box.
EffectPossiblecauseVerificationCorrectiveaction
Floorlocknotengaged.
Greenplungeronfloorlockpost
isvisible.
Activatefloorlock-greenplunger
shouldNOTbevisible.
Insufficientstaticfrictionbetween
floorlockfootbumperandfloor.
Floorsurfaceisslipperyand/or
foreignmembranebetween
bumperandfloor.
Cleanthesurroundingfloorand
removeloosematerialunderthe
bed.
Beddoesnotstayinplace
Floorlocknotfunctioning.
Floorlockisstuckinoneposition.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Bedisnotlevel
Hi/Loactuatorsareunevenly
extended
Beddeckisnotparallelwithfloor
Raisebedtohighestposition
or
Lowerbedtolowestposition
1153410-N~0059