EasyManuals Logo

Invacare Carroll CS3 User Manual

Invacare Carroll CS3
65 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Invacare®CarrollCSSeries™
EffectPossiblecauseVerificationCorrectiveaction
Actuatornotworking.SystemFault
NeedsReset
FlashingLEDsonACP,Bed
functionwillnotoperate.The
controlboxmaybeepwhen
pendantfunctionispressed.
ToReset-holdthependant
bedupandbeddownbuttons
for5+seconds.Thebedwill
beepandstopbeepingwhen
resetiscomplete.Thenpress
thebedupbutton.Ifthebed
beeps,theactuatorsareoutof
sync.Continuerisingtothe
highestpositiontosynchronize
theactuators.
Actuatornotworking.HI/LOlock-outison.Fornon-openbus(CS3,CS5)-A
Lockoutbox(asquareblackbox)
withatoggleswitch.Whenthe
switchistowards"ON"side,the
HILOonthehandpendantwill
notfunction.ThereisnoLED
light.
Foropenbusbeds(CS7,CS9,
CS9FX600)-Thelockoutbox(a
flatterrectangularwhitebox)with
atoggledomebuttonandLED
light.Whenthelightison,the
HILOonthehandpendantwill
notfunction.
Rotatethelockoutdialtothe
“unlocked”position(horizontal),
orflipswitchtoOFF.
LockoutLEDontheAttendant
ControlPad(ACP)isON.
Toggle-offtheindividualfunction
lockout(LEDshouldbeOFF)
Loosewireconnections.
Checkcableconnectionsvisually.Connectanyloosecablecouplings
and/orpowersupplies.
601153410-N~00

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Carroll CS3 and is the answer not in the manual?

Invacare Carroll CS3 Specifications

General IconGeneral
BrandInvacare
ModelCarroll CS3
CategoryMedical Equipment
LanguageEnglish

Related product manuals