Troubleshooting
EffectPossiblecauseVerificationCorrectiveaction
Faultyactuator.Disconnectpowersupply,swap
cablefromunresponsiveactuator
withafunctionalactuator.
Reconnectpowerandtest
functions.Faultyactuatorwillnot
workregardlessoftheconnection
port.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Actuatornotworking.
Faultyhandcontrolpendant.Pendantcableisconnected,power
ison,lock-outisoffandhand
controlpendantdoesnotfunction.
Disconnectpowersupplythen
replacehandpendantwith
functioningunit(fromanother
bed,ifavailable).Reconnect
powerandtest.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Bedendstonotfitproperly.
Bentbrackets.Visuallyensurethebracketsare
paralleltotheframe.
Carefullystraightenminor
deformations.CallService
Departmenttoreplace
significantlybentbrackets,
1-800-668-2337.
1153410-N~0061