Page #61 background imageLoading...
Page #61 background image
Invacare®CarrollCSSeries™
EffectPossiblecauseVerificationCorrectiveaction
Actuatornotworking.SystemFault
NeedsReset
FlashingLEDsonACP,Bed
functionwillnotoperate.The
controlboxmaybeepwhen
pendantfunctionispressed.
ToReset-holdthependant
bedupandbeddownbuttons
for5+seconds.Thebedwill
beepandstopbeepingwhen
resetiscomplete.Thenpress
thebedupbutton.Ifthebed
beeps,theactuatorsareoutof
sync.Continuerisingtothe
highestpositiontosynchronize
theactuators.
Actuatornotworking.HI/LOlock-outison.Fornon-openbus(CS3,CS5)-A
Lockoutbox(asquareblackbox)
withatoggleswitch.Whenthe
switchistowards"ON"side,the
HILOonthehandpendantwill
notfunction.ThereisnoLED
light.
Foropenbusbeds(CS7,CS9,
CS9FX600)-Thelockoutbox(a
flatterrectangularwhitebox)with
atoggledomebuttonandLED
light.Whenthelightison,the
HILOonthehandpendantwill
notfunction.
Rotatethelockoutdialtothe
“unlocked”position(horizontal),
orflipswitchtoOFF.
LockoutLEDontheAttendant
ControlPad(ACP)isON.
Toggle-offtheindividualfunction
lockout(LEDshouldbeOFF)
Loosewireconnections.
Checkcableconnectionsvisually.Connectanyloosecablecouplings
and/orpowersupplies.
601153410-N~00