Troubleshooting
FlashcodeFault
Consequenceforthe
scooterComments
5
BrakefailureStopsdrivingEnsurethatthedisengagingleverisintheengaged
position.
Thereisadefectinthebrakingcoilorinthecabling.
Checkthemagneticbrakeandcablingforopen
orshort-circuitedcircuitry.ContactyourInvacare
provider .
6
Noneutralpositionwhen
switchingscooteron.
StopsdrivingDriveleverisnotinneutralwhenthekeyswitchwas
turned.Putthedriveleverinneutral,turnthepower
offandthenturnonagain.
Itmaybenecessarytoreplacethedrivelever.Contact
yourInvacareprovider.
7
Faultinspeed
potentiometer
StopsdrivingThedrivelevercontrolscouldbefaultyor
incorrectlyconnected.Checkthecablingforopenor
short-circuitedcircuitry.
Potentiometerisnotcorrectlyadjustedandmustbe
replaced.ContactyourInvacareprovider .
8
MotorvoltageerrorStopsdrivingThemotororitscablingisdefective.Checkthecabling
foropenorshort-circuitedcircuitry.
9
Miscellaneousinternal
fault
StopsdrivingContactyourInvacareprovider.
10
Push/freewheelmode
error
Stopsmoving
Thescooterhasexceededthepermissiblemaximum
speedduringpushingorfreewheeling.Switchthe
electronicssystemoffandonagain.