Troubleshooting
FlashcodeFault
Consequenceforthe
scooterComments
5
BrakefailureStopsdriving•Ensurethatthedisengagingleverisintheengaged
position.
•Thereisadefectinthebrakingcoilorinthecabling.
Checkthemagneticbrakeandcablingforopen
orshort-circuitedcircuitry.ContactyourInvacare
provider .
6
Noneutralpositionwhen
switchingscooteron.
Stopsdriving•Driveleverisnotinneutralwhenthekeyswitchwas
turned.Putthedriveleverinneutral,turnthepower
offandthenturnonagain.
•Itmaybenecessarytoreplacethedrivelever.Contact
yourInvacareprovider.
7
Faultinspeed
potentiometer
Stopsdriving•Thedrivelevercontrolscouldbefaultyor
incorrectlyconnected.Checkthecablingforopenor
short-circuitedcircuitry.
•Potentiometerisnotcorrectlyadjustedandmustbe
replaced.ContactyourInvacareprovider .
8
MotorvoltageerrorStopsdriving•Themotororitscablingisdefective.Checkthecabling
foropenorshort-circuitedcircuitry.
9
Miscellaneousinternal
fault
Stopsdriving•ContactyourInvacareprovider.
10
Push/freewheelmode
error
Stopsmoving
•Thescooterhasexceededthepermissiblemaximum
speedduringpushingorfreewheeling.Switchthe
electronicssystemoffandonagain.
1576371-C45