Digital Intra-oral X-Ray Imaging System Pluto0001X /Pluto0002X User's Manual
1.2 Notes for Using
Otherwise, problems may occur and the equipment may not
function correctly.
Before using
Please check whether the USB connector is dry or clean
before connecting the USB connector
Please hold the control box of the USB when plugging the
USB connector, do not touch the pin of USB connector
During using
Do not move the USB connector during the use of the
sensor
will increase. Please pay attention to the temperature of the
The detector should warm up for 15 minutes before
exposure or updating the gain map or defect map.
During exposure
Do not move the Cable or Sensor during exposure, or it
may cause image noise or artifacts, even incorrect images.
Do not use the devices near the equipment generating a
strong magnetic field. Otherwise, it may cause image noise,
artifacts or even incorrect images.
After using
After the USB port is pulled out, please take care of the
USB connector to avoid the risk of damage.
The sensor should be stored in a place free of chemicals
or gases and free from adverse factors such as pressure, high
into the product package box, to avoid damage.
1.3 Storage and Operation Environment