Digital Intra-oral X-Ray Imaging System Pluto0001X /Pluto0002X User's Manual
AcqDelay 50 ms
Image cycle time 6 s
Trigger Mode
One is be triggered by the sensor automatically, the other
is trigged by the software command by an operator. Software
trigger mode is only used for debug and can be used by service
engineer who is authorized by iRay.
Type DC X-tube AC X-tube
mechanism Automatically detect
x-ray start and x-ray end
Only detect x-ray start;
x-ray end is pre-set by integration
time
Integration time Depend on X-ray width
AED(auto exposure detection) mode-DC x-ray tube
AED(auto exposure detection) mode-AC x-ray tube