Digital Intra-oral X-Ray Imaging System Pluto0001X /Pluto0002X User's Manual
5.2.2 EMS Compliance Table
Enclosure USB Port
Phenomenon Basic EMC standard Immunity test levels
Professional healthcare
facility environment
Electrostatic
Discharge
IEC 61000-4-2 ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV
air
IEC 61000-4-3 3V/m
80MHz-2.7GHz
80% AM at 1kHz
wireless communications
equipment
IEC 61000-4-3 Refer to table “Near
communications equipment”
Rated power frequency
IEC 61000-4-8 30A/m
50Hz or 60Hz
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Recommended separation distances between portable or