34
Digital Intra-oral X-Ray Imaging System Pluto0001X /Pluto0002X User's Manual
Incisors Child 70kV/8mA 0.025
Adult 70kV/8mA 0.032
Premolars and
Canines
Child 70kV/8mA 0.032
Adult 70kV/8mA 0.040
Molars Child 70kV/8mA 0.040
Adult 70kV/8mA 0.050
Note: X-ray sources: 70kV/8mA when using a 20cm SID
Recommended exposure time can vary depending on the
patient’s body size, age, sex, and thickness of the position to be
photographed.
Forlarger body types: increase the source current by25%
For children(5~21age): reduce the source current (or
Exposure time) by20%
For edentulous patients : reduce the source current by20%.
Since the X-ray exposure condition can be changed depending
on the age, gender and bone density of the patient, in case of
Pediatric, X-ray exposure condition can be changed by expert’s
For further information, please refer to FDA Pediatric
X-ray Imaging webpage, http://www.fda.gov/radiation-
emittingproducts/radiationemittingproductsandprocedures/
medicalimaging/ucm298899.htm)