EasyManuals Logo

ITech IT6015C-80-450S User Manual

ITech IT6015C-80-450S
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
Power Supply Function
start
voltage
Set the initial output voltage of the in-
strument before the LV123 regulation
waveform test. At the same time, this
value is also the end voltage after the
regulatory waveform test is
completed.
start
time
Test time of starting voltage, range:
0~999.999S.
end time Test time of end voltage, range:
0~999.999S.
count Set the number of test repetitions,
ranging from 1 to 65535. The actual
number of test executions = count *
3.
Save To
File
Set the save address of the waveform
file, range: 1~10.
User-
de-
fined
Select the user-defined LV123 waveform.
Run Run the presently recalled waveform.
Open Recall the waveform saved inside the
instrument.
Recall
File
The address of the waveform file,
ranging from 1 to 10.
Edit Edit a user-defined waveform.
unlimit-
ed
Range of unlimited operating capabil-
ity, with the following settings:
V1: Voltage in interval 1
V2: Voltage in interval 2
V3: Voltage in interval 3
V4: Voltage in interval 4
upper-
limited
Range of upper limited operating ca-
pability, with the following settings:
V1: Voltage in interval 1
V2: Voltage in interval 2
V3: Voltage in interval 3
Copyright © Itech Electronic Co., Ltd.
92

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT6015C-80-450S and is the answer not in the manual?

ITech IT6015C-80-450S Specifications

General IconGeneral
BrandITech
ModelIT6015C-80-450S
CategoryPower Supply
LanguageEnglish

Related product manuals