9. Product Specifications
9.1 Delivery form
Programat EP 5010
Power cord
Vacuum hose
Programat Firing Tray Kit 2
Automatic Temperature Checking Set ATK2 (test set)
USB download cable
Cooling grid
Programat WLAN Kit
DSA Holder Kit
Various accessories
Recommended accessories
Automatic Temperature Checking Set (ATK2)
Vacuum pump VP5
9.2 Technical data
Power supply 110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
Overvoltage category II
Pollution degree 2
Tolerated voltage fluctuations ± 10%
Max. power consumption 12 A at 110 120 V
8 A at 200 240 V
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used.
Values of electric fuses
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T15A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
Dimensions of electric fuses 110 120 V: Diameter 6.3 x 32 mm
200 240 V: Diameter 5 x 20 mm
Dimensions of the closed furnace Depth: 495 mm
Width: 320 mm / 395 mm (with cooling tray)
height: 550 mm
Usable size of the firing chamber Diameter: 90 mm
Height: 80 mm
Max. firing temperature 1200 °C
Weight 20.5 kg
Safety notes
The furnace complies with the following guidelines:
EN 61010-1:2010
IEC 61010-1:2010
UL 61010-1:2012/R:2015
CAN/CSA-C22.2 No. 61010-1:2012 + UPD No. 1:2015
EN 61010-2-010:2014
IEC 61010-2-010:2014
UL 61010-2-010:2015
CAN/CSA-C22.2 No. 61010-2-010:2015
Radio protection / electromagnetic compatibility: EMC tested