EasyManua.ls Logo

Ivoclar Vivadent Programat P200 - Product Specifications; Delivery Form; Technical Data; Acceptable Operating Conditions

Ivoclar Vivadent Programat P200
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
9. Product Specifications
This chapter contains all the relevant
product specifications:
9.1 Delivery form
1 Programat P200
1 Power cord
1 Vacuum hose
3 Extra fuses
1 Operating Instructions
1 Warranty card
1 Screwdriver
Programat firing cards
Recommended accessories
Vacuum pump VP3
Programat Accessory Assortment (tongs,
firing trays large and small, Temperature
Checking Set)
Colour
Standard colour: white (RAL 9016)
9.2 Technical data
Power supply
Single-phase alternating current
200–240 V / 50–60 Hz
110–120 V / 50–60 Hz
Tolerated voltage fluctuations +/– 10%
Power consumption
Furnace with pump
200–240 V max. 1800 W
110–120 V max. 1300 W
Vacuum quality
3 LEDs indicating 25 %, 50 %, and
100 % vacuum quality.
Acceptable data for pumps from other
manufacturers
Max. performance: 300 W
Final vacuum: approx. 25 mbar
Use only tested pumps
Electrical fuses
200–240 V:
T 6.3 A (heating circuit) (24)
T 315 mA (controls) (26)
T 3.15 A (vacuum pump) (25)
110-120 V:
T 15A (heating circuit) (24)
T 500 mA (controls) (26)
T 5A (pump) (25)
Dimensions of fuses
200–240 V = Diameter 5 x 20 mm
110–120 V = Diameter 6.3 x 32 mm
Dimensions of the closed furnace
Width/depth/height/ = 415 x 390 x 296 mm
Usable size of the firing chamber
Diameter 80 mm, height 38 mm
Max. firing temperature
1200 °C (2192 °F)
Weight
Furnace head: 3.5 kg
Furnace base: 10.0 kg
Furnace complete: 13.5 kg
Safety information
The P200 complies with the following
guidelines:
IEC 1010-1/EN 61010, Part 1
UL and cUL standards
Radio protection /
electromagnetic compatibility
EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range
+5 °C to +35 °C (+41 °F to +95 °F)
Acceptable humidity range
80 % maximum relative humidity for
temperatures up to 31 °C (87.8 ° F)
gradually decreasing to 50 % relative
humidity at 40 °C (104 °F); condensation
excluded.
Acceptable ambient pressure
500 mbar to 1060 mbar
The furnace is tested for use at altitudes of
up to 2000 m above sea level.
9.4 Acceptable transportation
and storage conditions
Acceptable temperature range
–20 to +50 °C (–4 °F to +122 °F)
Acceptable humidity range
80 % relative humidity.
Acceptable ambient pressure
500 mbar to 1060 mbar
Use only original packaging of the
Programat P200 together with the
respective foam material for shipping
purposes.